Loading…
Name ID Label Description Type Access Class
cutoff_code 40DA Tipo de corte baixo R/W Index 4096
D1 40DB D1 The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. R/W Float 8
D2 40DC D2 The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. R/W Float 8
D3 40DD D3 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. R/W Float 8
D4 40DE D4 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. R/W Float 8
K1 40DF K1 The tube period, in microseconds, corresponding to a density equal to D1. R/W Float 8
K2 40E0 K2 The tube period, in microseconds, corresponding to a density equal to D2. R/W Float 8
K3 40E1 K3 The tube period, in microseconds, corresponding to a density equal to D3. R/W Float 8
K4 40E2 K4 The tube period, in microseconds, corresponding to a density equal to D4. R/W Float 8
DCoef 40E3 Temp coeff The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. R/W Float 8