Loading…
Name ID Label Description Type Access Class
live_zero_vol_flow_hzfreq 424F Frequency Factor The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. R/W Float 512
live_zero_vol_flow_hzrate 4250 Rate Factor The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. R/W Float 512
live_zero_vol_flow_p_u 4251 Pulses/Unit This parameter allows you to scale the frequency output in terms of the number of pulses per unit of flow. R/W Float 512
live_zero_vol_flow_u_p 4252 Units/Pulse This parameter allows you to scale the frequency output in terms of the number of units of flow required to generate one pulse. R/W Float 512
field_ver_zero_digital_units 4569 Field Verification Zero Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. R/W Enum 32
field_ver_zero_digital_value 456A Field Verification Zero R Float 10
discrete_action_code 43D6 Discrete Action Selects an action to be assigned to a discrete. R/W Enum 4096
discrete_assignment_code 43D7 Discrete Assignment Assigns a discrete to an action. Select none for no discrete assignment. R/W Enum 4096
pc_index 43F3 Discrete Event num R/W Index 4096
pc1_setpointA 43D9 Setpoint A The first programmable value of a process variable in which the Discrete Events compares with. R/W Float 16
Name ID Label Description Type Access Class