Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
sp_volume | 4072 | Base Unit | A standard engineering unit that serves as the volume component in a special unit. | R/W | Enum | 32 |
sp_volume_time | 4073 | Base Time | A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. | R/W | Enum | 32 |
sp_volume_fact | 4074 | Conversion Factor | A programmable divider for the base volume unit in a special unit. | R/W | Float | 32 |
sp_volume_rate_text | 4075 | Flow Label | A 1 to 8 character description of a special unit of flow. | R/W | ASCII | 2048 |
sp_volume_total_text | 4076 | Total Label | A 1 to 8 character description of a special unit of volume, which is based on the special unit of flow to represent total. | R/W | ASCII | 2048 |
fixfreq | 4077 | Fixed freq | A value (in Hertz) that will be indicated on the frequency output for the purpose of a loop test. | R/W | Float | 4096 |
totctl | 4078 | Totlizer cntrl | Allows the user to Reset, Stop, or Start all totalizers. NOTE - Inventory totals are NOT reset. | R/W | Enum | 4096 |
fo_scaling_method | 401B | Frequency Output Scaling Method | This parameter allows you to scale the frequency output by using freq=flow, pulses/unit, or units/pulse. This parameter must be sent to the device before you make scaling changes. | R/W | Enum | 512 |
fo_polarity | 41D1 | Frequency Output Polarity | This parameter allows you to set the polarity of the frequency output. | R/W | Enum | 512 |
dens_tgco | 407A | DTG | This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |