Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
zero_pres_factor | 418A | Flowcal pressure | Pressure at which flow calibration was performed. | R/W | Float | 8 |
enable_pres_comp | 41D2 | Enable Pressure Comp | This parameter allows you to enable/disable the pressure correction feature | R/W | Enum | 8 |
static_pressure | 41D3 | Static Pressure | This parameter allows you to enter a static pressure when using the pressure correction feature | R/W | Float | 8 |
enable_extern_temp | 41E4 | Enable Ext Temp | This parameter allows you to enable/disable the external temperature feature | R/W | Enum | 8 |
static_temperature | 41E5 | Static Temperature | This parameter allows you to enter a static temperature when using the external temperature feature | R/W | Float | 8 |
flow_switch_setpoint | 418B | Flow Switch Setpoint | This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. | R/W | Float | 16 |
do1_assignment | 418C | DO is | This parameter identifies which condition the discrete output will act upon. | R/W | Enum | 1024 |
do1_state | 41CD | DO is | This parameter identifies the state of the discrete output. | R/W | Enum | 1024 |
do_index | 41B3 | DO index | R/W | Index | 4096 | |
chan_a | 418D | Channel A | This parameter identifies the physical output channel to be configured. Channel 1 is terminals 1 & 2, Channel 2 is terminals 3 & 4 and Channel 3 is terminals 5 & 6. | R | Enum | 64 |