Loading…
Name ID Label Description Type Access Class
zero_pres_factor 418A Flowcal pressure Pressure at which flow calibration was performed. R/W Float 8
enable_pres_comp 41D2 Enable Pressure Comp This parameter allows you to enable/disable the pressure correction feature R/W Enum 8
static_pressure 41D3 Static Pressure This parameter allows you to enter a static pressure when using the pressure correction feature R/W Float 8
enable_extern_temp 41E4 Enable Ext Temp This parameter allows you to enable/disable the external temperature feature R/W Enum 8
static_temperature 41E5 Static Temperature This parameter allows you to enter a static temperature when using the external temperature feature R/W Float 8
flow_switch_setpoint 418B Flow Switch Setpoint This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. R/W Float 16
do1_assignment 418C DO is This parameter identifies which condition the discrete output will act upon. R/W Enum 1024
do1_state 41CD DO is This parameter identifies the state of the discrete output. R/W Enum 1024
do_index 41B3 DO index R/W Index 4096
chan_a 418D Channel A This parameter identifies the physical output channel to be configured. Channel 1 is terminals 1 & 2, Channel 2 is terminals 3 & 4 and Channel 3 is terminals 5 & 6. R Enum 64
Name ID Label Description Type Access Class