Loading…
Name ID Label Description Type Access Class
xmtr_specific_status_7 0FBC Status group 16 Field Device Specific Status Byte 16- Data that has Field Device status information relating to: failures / warnings / status of process's. R BitEnum 2051
operating_mode_1 0FC6 Operate mode 1 Field Device Specific Operating Mode 1- A possible mode of operation in Field Device R Enum 2049
operating_mode_2 0FC7 Operate mode 2 Field Device Specific Operating Mode 2- A possible mode of operation in Field Device R Enum 2049
analog_output_saturated1 0FC8 AO saturated Field Device Specific Dynamic Variable Analog Output Saturated- Data that indicates which of the Analog Outputs are, or are not, saturated. The analog and digital Analog Outputs for the Dynamic Variable are beyond their limits and no longer represent the true applied process. R BitEnum 67
analog_output_saturated2 0FC9 AO saturated Field Device Specific Dynamic Variable Analog Output Saturated- Data that indicates which of the Analog Outputs are, or are not, saturated. The analog and digital Analog Outputs for the Dynamic Variable are beyond their limits and no longer represent the true applied process. R BitEnum 67
analog_output_saturated3 0FCA AO saturated Field Device Specific Dynamic Variable Analog Output Saturated- Data that indicates which of the Analog Outputs are, or are not, saturated. The analog and digital Analog Outputs for the Dynamic Variable are beyond their limits and no longer represent the true applied process. R BitEnum 67
analog_output_fixed1 0FCB AO fixed Field Device Specific Dynamic Variable Analog Output Fixed- Data that indicates which of the Analog Outputs are, or are not, Fixed. The analog and digital Analog Outputs for the Dynamic Variable are held at a constant value (depending on the Mode of the Field Device). They will not respond to the applied process. R BitEnum 67
analog_output_fixed2 0FCC AO fixed Field Device Specific Dynamic Variable Analog Output Fixed- Data that indicates which of the Analog Outputs are, or are not, Fixed. The analog and digital Analog Outputs for the Dynamic Variable are held at a constant value (depending on the Mode of the Field Device). They will not respond to the applied process. R BitEnum 67
analog_output_fixed3 0FCD AO fixed Field Device Specific Dynamic Variable Analog Output Fixed- Data that indicates which of the Analog Outputs are, or are not, Fixed. The analog and digital Analog Outputs for the Dynamic Variable are held at a constant value (depending on the Mode of the Field Device). They will not respond to the applied process. R BitEnum 67
primary_variable_code 0FD0 PV is Primary Index into Dynamic Variables Array- An index location, that indicates which Field Device dynamic variable has been mapped into the Primary Variable dynamic variable R/W Index 128
Name ID Label Description Type Access Class