Loading…
Name ID Label Description Type Access Class
D1 4099 D1 The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. R/W Float 8
D2 409A D2 The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. R/W Float 8
D3 409B D3 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. R/W Float 8
D4 409C D4 The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. R/W Float 8
K1 409D K1 The tube period, in microseconds, corresponding to a density equal to D1. R/W Float 8
K2 409E K2 The tube period, in microseconds, corresponding to a density equal to D2. R/W Float 8
K3 409F K3 The tube period, in microseconds, corresponding to a density equal to D3. R/W Float 8
K4 40A0 K4 The tube period, in microseconds, corresponding to a density equal to D4. R/W Float 8
DCoef 40A1 Temp coeff The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. R/W Float 8
DT 40A2 DT The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. R/W Float 8
Name ID Label Description Type Access Class