名称 ID 标签 描述 类型 访问 分类
dig_comm_fault_code 4091 Comm Fault ind. This fault setting applies to Modbus on RS485. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) 读/写 枚举 4
ao_1_fault_level 408D mA Fault Value This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. 读/写 浮点 64
fo_fault_level 4090 FO Fault Value This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. 读/写 浮点 64
density_for_cal 4092 读/写 浮点 4096
point_code 4093 读/写 无符号 4096
densityfd_for_cal 4094 读/写 浮点 4096
temp_for_cal 4095 Ref temperature User-entered temperature for the purpose of performing an on-line temperature calibration. 读/写 浮点 4096
cutoff_code 4098 切除类型 读/写 索引 4096
D1 4099 D1 The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. 读/写 浮点 8
D2 409A D2 The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. 读/写 浮点 8