名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
D3 | 409B | D3 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. | 读/写 | 浮点 | 8 |
D4 | 409C | D4 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. | 读/写 | 浮点 | 8 |
K1 | 409D | K1 | The tube period, in microseconds, corresponding to a density equal to D1. | 读/写 | 浮点 | 8 |
K2 | 409E | K2 | The tube period, in microseconds, corresponding to a density equal to D2. | 读/写 | 浮点 | 8 |
K3 | 409F | K3 | The tube period, in microseconds, corresponding to a density equal to D3. | 读/写 | 浮点 | 8 |
K4 | 40A0 | K4 | The tube period, in microseconds, corresponding to a density equal to D4. | 读/写 | 浮点 | 8 |
DCoef | 40A1 | Temp coeff | The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. | 读/写 | 浮点 | 8 |
DT | 40A2 | DT | The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. | 读/写 | 浮点 | 8 |
alarm_number | 40B4 | Event num | 读/写 | 索引 | 4096 | |
alarm1_type | 40A8 | Event1 type | Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. | 读/写 | 枚举 | 1024 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |