Loading…
Name ID Label Description Type Access Class
tube_amplitude_L 4105 Left Pickoff Amplitude R Float 10
tube_amplitude_R 4106 Right Pickoff Amplitude R Float 10
drive_gain 44AB Drive The drive gain of the transmitter. R Float 10
live_zero 4107 Live Zero Flow Rate This flow value is not effected by the flow cutoff and is damped at 8 seconds. It is to be used for diagnostic purposes only and is not used in ANY flow calculations or totalization. R Float 10
board_temp_700 4108 Board Temperature The board temperature of the core processor. R Float 10
case_temp 410A Meter Temperature (T-Series) The meter temperature of the T-Series sensor. R Float 10
FD 40F9 FD A calibration constant generated during a flowing density calibration. R/W Float 8
rs485_protocol 410B Protocol This parameter defines the protocol that will be used on the RS485 physical layer. R/W Enum 2048
rs485_baudrate 410C Baud Rate This parameter defines the baud rate for the protocol that will be used on the RS485 physical layer. R/W Enum 2048
rs485_parity 410D Parity This parameter defines the parity for the protocol that will be used on the RS485 physical layer. R/W Enum 2048
Name ID Label Description Type Access Class