Name ID Label Description Type Access Class
dens_fqco2 40FE DFQ2 This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. R/W Float 8
flow_tgco 40FF FTG This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. R/W Float 8
flow_fqco 4100 FFQ This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. R/W Float 8
max_zero 40D6 Zero Time The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. R/W Unsigned 12
fo_mode 40D3 Mode This parameter allows you to set the mode of the second frequency output when dual frequency outputs are configured. R/W Enum 512
pulse_width 40D4 Maximum Pulse Width A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. R/W Float 512
tempcal 40EC Temperature Calibration Factor A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. R/W ASCII 32
rt_zero 40D7 Zero Value A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. R/W Float 10
zero_offset_units 40D8 Zero units Units for zero offset - always microseconds. R/W Unsigned 32
rt_sigma 40D9 Standard Deviation A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. R Float 10
Name ID Label Description Type Access Class