Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
densityfd_for_cal | 40F0 | R/W | Float | 4096 | ||
temp_for_cal | 40ED | Ref temperature | User-entered temperature for the purpose of performing an on-line temperature calibration. | R/W | Float | 4096 |
cutoff_code | 40EB | Тип выкл. | R/W | Index | 4096 | |
D1 | 40F1 | D1 | The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. | R/W | Float | 8 |
D2 | 40F2 | D2 | The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. | R/W | Float | 8 |
D3 | 40F3 | D3 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
D4 | 40F4 | D4 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
K1 | 40F5 | K1 | The tube period, in microseconds, corresponding to a density equal to D1. | R/W | Float | 8 |
K2 | 40F6 | K2 | The tube period, in microseconds, corresponding to a density equal to D2. | R/W | Float | 8 |
K3 | 40F7 | K3 | The tube period, in microseconds, corresponding to a density equal to D3. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |