Loading…
Name ID Label Description Type Access Class
fault_index 40DA Output Type R/W Index 4096
LMV_fault_timeout 40DC Fault Timeout This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. R/W Unsigned 12
ao_1_fault_code 40E0 Действие при сбое Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
ao_2_fault_code 40E8 Frequency Output Fault Action Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
fo_fault_code 40E4 Действие при сбое Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
dig_comm_fault_code 40E6 Действие при сбое This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
ao_1_fault_level 40E2 Уровень ошибки This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. R/W Float 64
fo_fault_level 40E5 Уровень ошибки This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. R/W Float 64
density_for_cal 40EE R/W Float 4096
point_code 40EF R/W Unsigned 4096
Name ID Label Description Type Access Class