Loading…
Name ID Label Description Type Access Class
mass_flow_lower_range_value 401E Mass flo LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
mass_flow_lower_range_value_1 401F Mass flo LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
mass_flow_digital_damping 4020 Flo damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 32
mass_flow_upper_sensor_limit 400F Mass flo URL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
mass_flow_lower_sensor_limit 4021 Mass flo LRL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
mass_flow_minimum_span 4022 Mass flo min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
mass_flow_hzfreq 4023 Freq factr The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. R/W Float 512
mass_flow_hzrate 4024 Rate factr The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. R/W Float 512
mass_flow_p_u 4026 Pulses/Unit This parameter allows you to scale the frequency output in terms of the number of pulses per unit of flow. R/W Float 512
mass_flow_u_p 4027 Units/Pulse This parameter allows you to scale the frequency output in terms of the number of units of flow required to generate one pulse. R/W Float 512
Name ID Label Description Type Access Class