Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
enable_pres_comp | 41BC | Enable Pressure Comp | This parameter allows you to enable/disable the pressure correction feature | R/W | Enum | 8 |
static_pressure | 41BD | Static Pressure | This parameter allows you to enter a static pressure when using the pressure correction feature | R/W | Float | 8 |
enable_extern_temp | 41CE | Enable Ext Temp | This parameter allows you to enable/disable the external temperature feature | R/W | Enum | 8 |
static_temperature | 41CF | Static Temperature | This parameter allows you to enter a static temperature when using the external temperature feature | R/W | Float | 8 |
flow_switch_setpoint | 4100 | Flow Switch Setpoint | This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. | R/W | Float | 16 |
do1_assignment | 41D0 | DO is | This parameter identifies which condition the discrete output will act upon. | R/W | Enum | 1024 |
do1_state | 41D1 | DO is | This parameter identifies the state of the discrete output. | R/W | Enum | 1024 |
do_index | 41D2 | DO index | R/W | Index | 4096 | |
board_temp_units | 41E4 | Temp unit | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. | R | Enum | 32 |
Process_Variable_Code_0 | 41ED | Var Code 0 | The process variable of which display precision can be set | R | Enum | 2048 |
Name | ID | Label | Description | Type | Access | Class |