Name ID Label Description Type Access Class
alarm_select 400F AO Alrm typ Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. R Enum 10
mmi_transfer_function 4010 Xfer fnctn R/W Enum 32
analog_units 4011 Anlog unit R/W Enum 32
mass_flow_upper_range_value 4012 Mass flo URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
mass_flow_lower_range_value 4013 Mass flo LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
mass_flow_digital_damping 4014 Flo damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 32
serial_number 4015 Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
mass_flow_upper_sensor_limit 4008 Mass flo URL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
mass_flow_lower_sensor_limit 4016 Mass flo LRL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
mass_flow_minimum_span 4017 Mass flo min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
Name ID Label Description Type Access Class