Name ID Label Description Type Access Class
zero_offset_units 4099 Zero units Units for zero offset - always microseconds. R/W Unsigned 32
rt_sigma 409A Std dev A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. R Float 10
fault_index 409B Output Type R/W Index 4096
LMV_fault_timeout 409D Fault Timeout This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. R/W Unsigned 12
ao_1_fault_code 40A1 AO1 Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
fo_fault_code 40A5 FO Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
dig_comm_fault_code 40A7 Comm Fault ind. This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
ao_1_fault_level 40A3 mA1 Fault Value This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. R/W Float 64
fo_fault_level 40A6 FO Fault Value This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. R/W Float 64
density_for_cal 40A8 R/W Float 4096
Name ID Label Description Type Access Class