Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
fo_polarity | 41BB | Polarity | This parameter allows you to set the polarity of the frequency output. | R/W | Enum | 512 |
dens_tgco | 4090 | DTG | This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. | R/W | Float | 8 |
dens_fqco1 | 4091 | DFQ1 | This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. | R/W | Float | 8 |
dens_fqco2 | 4092 | DFQ2 | This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_tgco | 4093 | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_fqco | 4094 | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | R/W | Float | 8 |
max_zero | 4095 | Zero time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | R/W | Unsigned | 12 |
pulse_width | 4096 | Max pulse width | A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. | R/W | Float | 512 |
tempcal | 4097 | Temp cal factr | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | R/W | ASCII | 32 |
rt_zero | 4098 | Zero | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | R/W | Float | 10 |
Name | ID | Label | Description | Type | Access | Class |