Name ID Label Description Type Access Class
analog_units 4011 Anlog unit R/W Enum 32
mass_flow_upper_range_value 4012 Mass flo URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
mass_flow_lower_range_value 4013 Mass flo LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
mass_flow_digital_damping 4014 Flo damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 32
serial_number 4015 Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
mass_flow_upper_sensor_limit 4008 Mass flo URL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
mass_flow_lower_sensor_limit 4016 Mass flo LRL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
mass_flow_minimum_span 4017 Mass flo min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
mass_flow_hzfreq 4018 Freq factr The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. R/W Float 512
mass_flow_hzrate 401A Rate factr The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. R/W Float 512