Loading…
名称 ID 标签 描述 类型 访问 分类
ao_1_fault_code 40A1 AO1 Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. 读/写 枚举 4
fo_fault_code 40A5 FO Fault indicator Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. 读/写 枚举 4
dig_comm_fault_code 40A7 Comm Fault ind. This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) 读/写 枚举 4
ao_1_fault_level 40A3 mA1 Fault Value This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. 读/写 浮点 64
fo_fault_level 40A6 FO Fault Value This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. 读/写 浮点 64
density_for_cal 40A8 读/写 浮点 4096
point_code 40A9 读/写 无符号 4096
densityfd_for_cal 40AA 读/写 浮点 4096
temp_for_cal 40AB Ref temperature User-entered temperature for the purpose of performing an on-line temperature calibration. 读/写 浮点 4096
cutoff_code 40AE 切除类型 读/写 索引 4096