名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
flow_tgco | 4093 | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
flow_fqco | 4094 | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
max_zero | 4095 | Zero time | The number of readings collected from the sensor during an Auto Zero operation. | 读/写 | 无符号 | 4 |
pulse_width | 4096 | Max pulse width | A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. | 读/写 | 浮点 | 512 |
tempcal | 4097 | Temp cal factr | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | 读/写 | ASCII | 32 |
rt_zero | 4098 | 归零 | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | 读/写 | 浮点 | 2 |
zero_offset_units | 4099 | Zero units | Units for zero offset - always microseconds. | 读/写 | 无符号 | 32 |
rt_sigma | 409A | Std dev | A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. | 读 | 浮点 | 2 |
fault_index | 409B | Output Type | 读/写 | 索引 | 4096 | |
LMV_fault_timeout | 409D | Fault Timeout | This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. | 读/写 | 无符号 | 4 |