名称 ID 标签 描述 类型 访问 分类
flow_tgco 4093 FTG This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. 读/写 浮点 8
flow_fqco 4094 FFQ This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. 读/写 浮点 8
max_zero 4095 Zero time The number of readings collected from the sensor during an Auto Zero operation. 读/写 无符号 4
pulse_width 4096 Max pulse width A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. 读/写 浮点 512
tempcal 4097 Temp cal factr A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. 读/写 ASCII 32
rt_zero 4098 归零 A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. 读/写 浮点 2
zero_offset_units 4099 Zero units Units for zero offset - always microseconds. 读/写 无符号 32
rt_sigma 409A Std dev A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. 浮点 2
fault_index 409B Output Type 读/写 索引 4096
LMV_fault_timeout 409D Fault Timeout This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. 读/写 无符号 4