名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
fault_index | 4060 | Output Type | 读/写 | 索引 | 4096 | |
ao_1_fault_code | 46E3 | 故障动作 | Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. | 读/写 | 枚举 | 12 |
ao_2_fault_code | 46E9 | 故障动作 | Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. | 读/写 | 枚举 | 12 |
dig_comm_fault_code | 46FB | 故障动作 | This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) | 读/写 | 枚举 | 12 |
discrete_output_fault_code | 46FC | Fault Action | This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) | 读/写 | 枚举 | 12 |
ao_1_fault_level | 46E4 | 故障等级 | This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 3.2 to 3.6 mA for downscale or 21.0 to 21.5 for upscale. | 读/写 | 浮点 | 64 |
ao_2_fault_level | 46EA | 故障等级 | This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 3.2 to 3.6 mA for downscale or 21.0 to 21.5 for upscale. | 读/写 | 浮点 | 64 |
LMV_fault_timeout | 4064 | Fault Timeout | This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. | 读/写 | 无符号 | 12 |
history_0 | 4A3B | Alarm Event Log Index | 读 | 无符号 | 2048 | |
history_1 | 4A3C | Alarm Event Log Index | 读 | 无符号 | 2048 | |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |