Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
api_thermal_exp_coeff | 40AE | Thermal Expansion Coefficient | This parameter allows you to select the Thermal Expansion Coefficient API. The units are the inverse of the temperature user units. | R/W | Float | 8 |
api_ref_temp | 40AF | Reference Temperature | This parameter allows you to select the Reference Termperature for API. The units must be the same as the temperature user units. | R/W | Float | 8 |
enable_api | 40B2 | Enable API | This parameter allows you to enable/disable the api feature | R/W | Enum | 8 |
alternate_pressure | 40B3 | Reference Pressure | The pressure to which referred density data will be corrected. | R/W | Float | 8 |
api_density_K0 | 40B4 | API K0 | R/W | Float | 8 | |
api_density_K1 | 40B5 | API K1 | R/W | Float | 8 | |
api_ctl | 40B1 | CTL | This parameter displays the Correction of Temperature on a Liquid. | R | Float | 10 |
enable_extern_temp | 40B8 | 外部温度 | This parameter allows you to enable/disable the external temperature feature | R/W | Enum | 8 |
static_temperature | 40B9 | Fixed or Polled Temperature | This parameter allows you to enter a static temperature when using the external temperature feature | R/W | Float | 8 |
enable_extern_temp_modulus | 40BA | 外部温度モジュール | Enabled: Process measurement will be adjusted for the effect of temperature on the vibrating element. Disabled: Process measurement will not be adjusted for the effect of temperature. | R/W | Enum | 8 |