Name ID Label Description Type Access Class
fault_index 4060 Output Type R/W Index 4096
ao_1_fault_code 46E3 故障アクション Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
ao_2_fault_code 46E9 故障アクション Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
dig_comm_fault_code 46FB 故障アクション This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
discrete_output_fault_code 46FC Fault Action This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
ao_1_fault_level 46E4 故障レベル This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 3.2 to 3.6 mA for downscale or 21.0 to 21.5 for upscale. R/W Float 64
ao_2_fault_level 46EA 故障レベル This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 3.2 to 3.6 mA for downscale or 21.0 to 21.5 for upscale. R/W Float 64
LMV_fault_timeout 4064 Fault Timeout This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. R/W Unsigned 12
history_0 4A3B Alarm Event Log Index R Unsigned 2048
history_1 4A3C Alarm Event Log Index R Unsigned 2048
Name ID Label Description Type Access Class