Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
volume_flow_meter_factor | 4E5B | Volume Factor (0.8-1.2) | The adjustment to be applied to the volume flow process variable. The measured volume flow is multiplied by this value. | R/W | Float | 10 |
pressure_input_type | 4E5C | Pressure Input | R/W | Enum | 8 | |
base_dens_unit | 415C | 基準単位 | The unit from which the special unit will be calculated. | R/W | Enum | 32 |
dens_unit_conv_factor | 415D | 変換係数 | The value by which the base unit will be divided, to report data in the special unit. | R/W | Float | 32 |
sp_dens_unit_text | 415E | ラベル | The name to be used for the special unit. | R/W | ASCII | 2048 |
base_dynamic_visc_unit | 415F | 基準単位 | The unit from which the special unit will be calculated. | R/W | Enum | 32 |
dynamic_visc_unit_conv_factor | 4160 | 変換係数 | The value by which the base unit will be divided, to report data in the special unit. | R/W | Float | 32 |
sp_dynamic_visc_unit_text | 4161 | ラベル | The name to be used for the special unit. | R/W | ASCII | 2048 |
base_kinematic_visc_unit | 4162 | 基準単位 | The unit from which the special unit will be calculated. | R/W | Enum | 32 |
kinematic_visc_unit_conv_factor | 4163 | 変換係数 | The value by which the base unit will be divided, to report data in the special unit. | R/W | Float | 32 |
Name | ID | Label | Description | Type | Access | Class |