Name ID Label Description Type Access Class
edconcFamily 48BF ED Concentration Family Device Family- Indicates the family (i.e, pressure, temperature, flow, etc.) that the device variable belongs. R Enum 2048
edconcPDQ 4211 Status Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
edconcLS 48C0 ED Concentration LS Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
edconcStatus 48C1 ED Concentration Family Status Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 10
edconcUpdatePeriod 48C2 ED Concentration Update Period R Time 0
drive_digital_units 4824 Drive Gain Unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R Enum 32
drive_digital_value 4093 Drive Gain Drive Gain is the amount of current required to drive the vibrating element, expressed as a percentage of the maximum allowable drive gain. R Float 10
drive_upper_sensor_limit 429B Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
drive_lower_sensor_limit 4825 Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
drive_minimum_span 4826 Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8