Name ID Label Description Type Access Class
Last_Cleared 406C Last Cleared R Unsigned 4096
rt_zero 405B Текущ. нулевое знач. A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. R/W Float 10
rt_sigma 405D Standard Deviation A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. R Float 10
factory_zero_offset 405E Factory zero offset Factory zero offset value - always microseconds. R/W Float 10
fault_index 4060 Output Type R/W Index 4096
ao_1_fault_code 46E3 Действие при сбое Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
ao_2_fault_code 46E9 Действие при сбое Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. R/W Enum 12
dig_comm_fault_code 46FB Действие при сбое This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
discrete_output_fault_code 46FC Fault Action This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) R/W Enum 12
ao_1_fault_level 46E4 Уровень ошибки This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 3.2 to 3.6 mA for downscale or 21.0 to 21.5 for upscale. R/W Float 64
Name ID Label Description Type Access Class