Name ID Label Description Type Access Class
eddenssguPDQ 421B ED Density SGU PDQ Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
eddenssguLS 48AD ED Density SGU LS Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
eddenssguStatus 48AE ED Density SGU Family Status Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 10
eddenssguUpdatePeriod 48AF ED Density SGU Update Period R Time 0
mass_flow_digital_units 41E4 Mass Flow Rate Unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. R/W Enum 32
ed_net_mas_flow_digital_value 42CB Flow Rate R Float 10
mass_flow_digital_cutoff 485C Mass Flow Cutoff The mass flow rate below which the frequency/pulse output indicates zero flow, and below which totalizers stop counting. Cutoffs can also be configured for the mA output(s) (the AO cutoffs). An AO cutoff is active only if the process variable assigned to the output is mass flow or volume flow. R/W Float 8
mass_flow_digital_damping 48B0 Flow Damping Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 32
mass_flow_upper_sensor_limit 42CD Mass Flow Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
mass_flow_lower_sensor_limit 42CF Mass Flow Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
Name ID Label Description Type Access Class