Name ID Label Description Type Access Class
save_gas_cal_1 41F9 Save Gas Calibration 1 W Unsigned 2
save_gas_cal_2 41FA Save Gas Calibration 2 W Unsigned 2
save_gas_cal_3 41FB Save Gas Calibration 3 W Unsigned 2
start_gas_cal 41FC Start Gas Calibration W Unsigned 2
active_gas_cal_index 41CE Active Calibration The calibration factors to be used for measurement. R/W Index 2048
number_of_gas_cal_points 41CB Calibration Format Displays whether this calibration is based on two calibration gases or three calibration gases. R/W Enum 32
gas_stability 41CC Calibration Point Indicates whether or not the data for this calibration is acceptable. R Enum 2
sg_cal_stability 41CD Stability A measure of the variability of the Sensor Time Period value, used to indicate the stage or the quality of the calibration process. R/W Float 8
enable_external_volume_flow 41DF Volume Flow Source Enabled: Volume flow data will be obtained from an external device. Requires polling setup. Disabled: No volume flow data will be obtained. R/W Enum 32
external_volume_flow 41E0 External Flow The current volume flow rate from the external device. R/W Float 2
Name ID Label Description Type Access Class