Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
visc_trim_point | 40B6 | Visc trim point | R/W | Enum | 4096 | |
temp_for_cal | 40B7 | Ref temperature | User-entered temperature for the purpose of performing an on-line temperature calibration. | R/W | Float | 4096 |
cutoff_code | 40BB | Cutoff typ | R/W | Index | 4096 | |
D1 | 40BC | Dens A | The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. Dens A can represent: The Dens A (D1) value on an ELITE sensor, a value of 0.0000 g/cc for a D, DL, or DT sensor with a 14-character dens cal factor. | R/W | Float | 8 |
D2 | 40BD | Dens B | The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. Dens B can represent: The Dens B (D2) value on an ELITE sensor, a value of 1.0000 g/cc for a D, DL, or DT sensor with a 14-character dens cal factor. | R/W | Float | 8 |
K1 | 40BE | K1 | The tube period, in microseconds, corresponding to a density equal to Dens A. | R/W | Float | 8 |
K2 | 40BF | K2 | The tube period, in microseconds, corresponding to a density equal to Dens B. | R/W | Float | 8 |
DCoef | 40C0 | Temp coeff | The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. | R/W | Float | 8 |
alarm_number | 40D2 | Event num | R/W | Index | 4096 | |
alarm1_type | 40C6 | Event1 type | Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. | R/W | Enum | 1024 |