名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
zero_offset_units | 40AC | Zero units | Units for zero offset - always microseconds. | 读/写 | 无符号 | 32 |
rt_sigma | 40AD | Std dev | A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. | 读 | 浮点 | 2 |
fault_code | 40AE | Fault indicator | Downscale: An analog output goes to 0 or 2 mA, the frequency = 0 Hz. Upscale: Milliamp outputs = 22 mA, the frequency output = 15 kHz. Last measured value: Outputs = value measured immediately before before fault. Internal zero: Outputs tied to flow = zero flow. | 读/写 | 枚举 | 4 |
timeout | 40AF | Slug duration | A programmable time period, from 0.00 to 60.00 seconds, during which outputs representing flow hold their last measured value before indicating slug flow. | 读/写 | 浮点 | 4 |
densitya_for_cal | 40B0 | 参考密度a | 读/写 | 浮点 | 4096 | |
densityb_for_cal | 40B1 | 参考密度b | 读/写 | 浮点 | 4096 | |
poll_ctrl | 40B2 | 粘度轮询 | A variable which represents how [differential] pressure is brought into the transmitter for the purpose of viscosity calculation or pressure compensation. | 读/写 | 枚举 | 128 |
viscal | 40B3 | Visc cal factr | A 14-character factor consisting of a slope and offset (optional) used to adjust the measured viscosity. | 读/写 | ASCII | 32 |
dp_tag | 40B4 | 差压现场设备位号 | A tag used to identify a pressure transmitter being polled by the 9739 via a HART network. | 读/写 | 压缩文本 | 128 |
viscosity_for_cal | 40B5 | 参考粘度 | User-entered viscosity for the purpose of performing an on-line viscosity calibration. | 读/写 | 浮点 | 4096 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |