Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
zero_offset_units | 40AC | Zero units | Units for zero offset - always microseconds. | R/W | Unsigned | 32 |
rt_sigma | 40AD | Std dev | A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. | R | Float | 2 |
fault_code | 40AE | Fault indicator | Downscale: An analog output goes to 0 or 2 mA, the frequency = 0 Hz. Upscale: Milliamp outputs = 22 mA, the frequency output = 15 kHz. Last measured value: Outputs = value measured immediately before before fault. Internal zero: Outputs tied to flow = zero flow. | R/W | Enum | 4 |
timeout | 40AF | Slug duration | A programmable time period, from 0.00 to 60.00 seconds, during which outputs representing flow hold their last measured value before indicating slug flow. | R/W | Float | 4 |
densitya_for_cal | 40B0 | Densidade de referência a | R/W | Float | 4096 | |
densityb_for_cal | 40B1 | Densidade de referência b | R/W | Float | 4096 | |
poll_ctrl | 40B2 | Pesquisa de viscosidade | A variable which represents how [differential] pressure is brought into the transmitter for the purpose of viscosity calculation or pressure compensation. | R/W | Enum | 128 |
viscal | 40B3 | Visc cal factr | A 14-character factor consisting of a slope and offset (optional) used to adjust the measured viscosity. | R/W | ASCII | 32 |
dp_tag | 40B4 | Tag de dispositivo de campo DP | A tag used to identify a pressure transmitter being polled by the 9739 via a HART network. | R/W | PackedAscii | 128 |
viscosity_for_cal | 40B5 | Viscosidade de referência | User-entered viscosity for the purpose of performing an on-line viscosity calibration. | R/W | Float | 4096 |
Name | ID | Label | Description | Type | Access | Class |