Loading…
Name ID Label Description Type Access Class
zero_offset_units 40AC Zero units Units for zero offset - always microseconds. R/W Unsigned 32
rt_sigma 40AD Std dev A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. R Float 2
fault_code 40AE Fault indicator Downscale: An analog output goes to 0 or 2 mA, the frequency = 0 Hz. Upscale: Milliamp outputs = 22 mA, the frequency output = 15 kHz. Last measured value: Outputs = value measured immediately before before fault. Internal zero: Outputs tied to flow = zero flow. R/W Enum 4
timeout 40AF Slug duration A programmable time period, from 0.00 to 60.00 seconds, during which outputs representing flow hold their last measured value before indicating slug flow. R/W Float 4
densitya_for_cal 40B0 Ref dens a R/W Float 4096
densityb_for_cal 40B1 Ref dens b R/W Float 4096
poll_ctrl 40B2 Viscosity polling A variable which represents how [differential] pressure is brought into the transmitter for the purpose of viscosity calculation or pressure compensation. R/W Enum 128
viscal 40B3 Visc cal factr A 14-character factor consisting of a slope and offset (optional) used to adjust the measured viscosity. R/W ASCII 32
dp_tag 40B4 DP fld dev tag A tag used to identify a pressure transmitter being polled by the 9739 via a HART network. R/W PackedAscii 128
viscosity_for_cal 40B5 Ref viscosity User-entered viscosity for the purpose of performing an on-line viscosity calibration. R/W Float 4096
Name ID Label Description Type Access Class