Name ID Label Description Type Access Class
visc_trim_point 40B6 Visc trim point R/W Enum 4096
temp_for_cal 40B7 Ref temperature User-entered temperature for the purpose of performing an on-line temperature calibration. R/W Float 4096
cutoff_code 40BB Cutoff typ R/W Index 4096
D1 40BC Dens A The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. Dens A can represent: The Dens A (D1) value on an ELITE sensor, a value of 0.0000 g/cc for a D, DL, or DT sensor with a 14-character dens cal factor. R/W Float 8
D2 40BD Dens B The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. Dens B can represent: The Dens B (D2) value on an ELITE sensor, a value of 1.0000 g/cc for a D, DL, or DT sensor with a 14-character dens cal factor. R/W Float 8
K1 40BE K1 The tube period, in microseconds, corresponding to a density equal to Dens A. R/W Float 8
K2 40BF K2 The tube period, in microseconds, corresponding to a density equal to Dens B. R/W Float 8
DCoef 40C0 Temp coeff The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. R/W Float 8
alarm_number 40D2 Event num R/W Index 4096
alarm1_type 40C6 Event1 type Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. R/W Enum 1024
Name ID Label Description Type Access Class