Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
hz_freq | 4022 | Коэф. частоты | The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. | R/W | Float | 512 |
hz_rate | 4023 | Коэф. скорости | The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. | R/W | Float | 512 |
slughi | 4024 | Slug high limit | A programmable high limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. | R/W | Float | 8 |
sluglo | 4025 | Slug low limit | A programmable low limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. | R/W | Float | 8 |
flocal | 4026 | FlowCal | A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. | R/W | ASCII | 32 |
fixfreq | 4027 | Фикс. част | A value (in Hertz) that will be indicated on the frequency output for the purpose of a loop test. | R/W | Float | 4096 |
rt_freq | 4028 | Pres freq | The frequency, in Hertz, presently being indicated on the frequency output. | R | Float | 514 |
totctl | 4029 | Управление сумматором | Allows the user to Reset all totalizers. NOTE - Inventory totals are NOT reset. | R/W | Enum | 4096 |
dencal | 402A | Коэф. калиб. плотн. | 14 characters describing the sensitivity of the sensor to density. | R/W | ASCII | 32 |
rt_zero | 402B | Zero value | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | R | Float | 2 |
Name | ID | Label | Description | Type | Access | Class |