Name ID Label Description Type Access Class
mass_flow_upper_sensor_limit 400B Mass Flow Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
mass_flow_lower_sensor_limit 401E Mass Flow Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
mass_flow_minimum_span 401F Mass Flow Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
mass_flow_hzfreq 4020 Frequency Factor The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. R/W Float 512
mass_flow_hzrate 4022 Rate Factor The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. R/W Float 512
mass_flow_p_u 4023 Pulses/Unit This parameter allows you to scale the frequency output in terms of the number of pulses per unit of flow. R/W Float 512
mass_flow_u_p 4025 Units/Pulse This parameter allows you to scale the frequency output in terms of the number of units of flow required to generate one pulse. R/W Float 512
massflowClassificationCode 4A6E Mass Flow Classification Classification- Indicates the unit code expansion table used for a particular dynamic or device variable. R Enum 0
massflowFamily 4A6F Mass Flow Family Device Family- Indicates the family (i.e, pressure, temperature, flow, etc.) that the device variable belongs. R Enum 2048
massflowPDQ 49E5 Mass Flow PDQ Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
Name ID Label Description Type Access Class