Loading…
名称 ID 标签 描述 类型 访问 分类
pv_disp12 4125 Position 12 This parameter allows you to select the process variables that will appear on the display. Up to 15 process variables can be configured to appear on the display. They can be accessed by using the scroll button. 读/写 枚举 256
pv_disp13 4126 Position 13 This parameter allows you to select the process variables that will appear on the display. Up to 15 process variables can be configured to appear on the display. They can be accessed by using the scroll button. 读/写 枚举 256
pv_disp14 4127 Position 14 This parameter allows you to select the process variables that will appear on the display. Up to 15 process variables can be configured to appear on the display. They can be accessed by using the scroll button. 读/写 枚举 256
pv_disp15 4128 Position 15 This parameter allows you to select the process variables that will appear on the display. Up to 15 process variables can be configured to appear on the display. They can be accessed by using the scroll button. 读/写 枚举 256
tot_startstop_enable 4129 Start/Stop Totalizers Enable or disable the ability to start and stop the totalizers from the local display. 读/写 枚举 256
Update_Period 412A 刷新率 (100 - 10,000) The period in which the process variables on the local display are updated. (in msec) 读/写 无符号 256
Backlight_Control 412B Control The state of the backlight on the local display. 读/写 枚举 256
Backlight_Intensity 412C Intensity (0-63) The amount of intensity of the backlight on the local display. Range is 0 to 63 where 0 is off and 63 is full on. 读/写 无符号 256
sens_type_code 40BA Type This parameter indicates which type of sensor is being used for process measurement. 读/写 枚举 8
output_opt_brd 412D Option Board This parameter represents the various types of output option boards that are available. 枚举 2048
名称 ID 标签 描述 类型 访问 分类