Loading…
Loading…
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
flow_tgco | 40C9 | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
flow_fqco | 40CA | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
max_zero | 40CB | Zero Time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | 读/写 | 无符号 | 12 |
pulse_width | 40CC | Maximum Pulse Width | A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. | 读/写 | 浮点 | 512 |
tempcal | 40CD | Temperature Calibration Factor | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | 读/写 | ASCII | 32 |
rt_zero | 40CE | Zero Value | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | 读/写 | 浮点 | 10 |
rt_sigma | 40CF | Standard Deviation | 标准偏差是从最后一次调零期间采集的调零样品数量获得的 2 西格玛级结果。“标准偏差”、“实时零点”、“智能零点校验”和“现场校验零点”可用于查看您仪表上的零点稳定程度。 | 读 | 浮点 | 10 |
fault_index | 40D0 | Output Type | 读/写 | 索引 | 4096 | |
LMV_fault_timeout | 40D2 | Fault Timeout | This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. | 读/写 | 无符号 | 12 |
ao_1_fault_code | 40D8 | 故障动作 | Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. | 读/写 | 枚举 | 12 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |