名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
met_direct | 40B5 | Flow Direction | This parameter enables the user to determine whether the outputs and totalizers will indicate forward, reverse, bi-directional or absolute value flow. | 读/写 | 枚举 | 32 |
slughi | 40B6 | 两相流上限 | 可编程密度上限,让变送器能够在流体密度超出设置的双相流限制时指示双相流(气团或液流中固体积聚引起的极端流体密度)。 | 读/写 | 浮点 | 8 |
sluglo | 40B7 | 两相流下限 | 可编程的密度下限,让变送器能够在流体密度超出设置的双相流限制时指示双相流(气团或液流中固体积聚引起的极端流体密度)。 | 读/写 | 浮点 | 8 |
slug_duration | 40B8 | 双相流持续时间 | 0.00 至 60.00 秒可编程时间段,在此期间代表流量的输出在指示双相流之前保持其最后测量的值。 | 读/写 | 浮点 | 12 |
flocal | 40B9 | FlowCal | A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. | 读/写 | ASCII | 32 |
fcf | 40BB | Flow FCF | A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. | 读/写 | ASCII | 32 |
ma_1_cutoff | 4028 | mA Output Flow Rate Cutoff | The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. | 读/写 | 浮点 | 8 |
ma_2_cutoff | 40BC | SV AO cutoff | The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. | 读/写 | 浮点 | 8 |
sp_volume | 40BD | 基础单位 | A standard engineering unit that serves as the volume component in a special unit. | 读/写 | 枚举 | 32 |
sp_volume_time | 40BE | 基础时间 | A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. | 读/写 | 枚举 | 32 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |