Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
flow_tgco | 40C9 | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | R/W | Float | 8 |
flow_fqco | 40CA | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | R/W | Float | 8 |
max_zero | 40CB | Zero Time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | R/W | Unsigned | 12 |
pulse_width | 40CC | Maximum Pulse Width | A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. | R/W | Float | 512 |
tempcal | 40CD | Temperature Calibration Factor | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | R/W | ASCII | 32 |
rt_zero | 40CE | Zero Value | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | R/W | Float | 10 |
rt_sigma | 40CF | Standard Deviation | Standard deviation is the two sigma result from the number of zero samples taken during the last zero. 'Standard Deviation', 'Live Zero', 'Smart Zero Verification' and 'Field Verification Zero' can be used to see how stable the zero is on your meter. | R | Float | 10 |
fault_index | 40D0 | Output Type | R/W | Index | 4096 | |
LMV_fault_timeout | 40D2 | Fault Timeout | This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. | R/W | Unsigned | 12 |
ao_1_fault_code | 40D8 | Fault Action | Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. | R/W | Enum | 12 |
Name | ID | Label | Description | Type | Access | Class |