Loading…
Name ID Label Description Type Access Class
met_direct 40B5 Flow Direction This parameter enables the user to determine whether the outputs and totalizers will indicate forward, reverse, bi-directional or absolute value flow. R/W Enum 32
slughi 40B6 Two-Phase High Limit A programmable high limit on density,which enables the transmitter to indicate two-phase flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed two-phase flow limits. R/W Float 8
sluglo 40B7 Two-Phase Low Limit A programmable low limit on density, which enables the transmitter to indicate two-phase flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed two-phase flow limits. R/W Float 8
slug_duration 40B8 Two-Phase Duration A programmable time period,from 0.00 to 60.00 seconds,during which outputs representing flow hold their last measured value before indicating two-phase flow. R/W Float 12
flocal 40B9 FlowCal A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. R/W ASCII 32
fcf 40BB Flow FCF A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. R/W ASCII 32
ma_1_cutoff 4028 mA Output Flow Rate Cutoff The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. R/W Float 8
ma_2_cutoff 40BC SV AO cutoff The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. R/W Float 8
sp_volume 40BD Base Unit A standard engineering unit that serves as the volume component in a special unit. R/W Enum 32
sp_volume_time 40BE Base Time A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. R/W Enum 32
Name ID Label Description Type Access Class