Name ID Label Description Type Access Class
dens_fqco1 40C7 DFQ1 This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. R/W Float 8
dens_fqco2 40C8 DFQ2 This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. R/W Float 8
flow_tgco 40C9 FTG This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. R/W Float 8
flow_fqco 40CA FFQ This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. R/W Float 8
max_zero 40CB Zero Time The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. R/W Unsigned 12
pulse_width 40CC Maximum Pulse Width A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. R/W Float 512
tempcal 40CD Temperature Calibration Factor A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. R/W ASCII 32
rt_zero 40CE Zero Value A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. R/W Float 10
rt_sigma 40CF Standard Deviation A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. R Float 10
fault_index 40D0 Output Type R/W Index 4096
Name ID Label Description Type Access Class