Name ID Label Description Type Access Class
pv_disp15 4128 Position 15 This parameter allows you to select the process variables that will appear on the display. Up to 15 process variables can be configured to appear on the display. They can be accessed by using the scroll button. R/W Enum 256
tot_startstop_enable 4129 Start/Stop Totalizers Enable or disable the ability to start and stop the totalizers from the local display. R/W Enum 256
Update_Period 412A Update Period (100-10,000) The period in which the process variables on the local display are updated. (in msec) R/W Unsigned 256
Backlight_Control 412B Control The state of the backlight on the local display. R/W Enum 256
Backlight_Intensity 412C Intensity (0-63) The amount of intensity of the backlight on the local display. Range is 0 to 63 where 0 is off and 63 is full on. R/W Unsigned 256
sens_type_code 40BA Type This parameter indicates which type of sensor is being used for process measurement. R/W Enum 8
output_opt_brd 412D Option Board This parameter represents the various types of output option boards that are available. R Enum 2048
poll_ctrl_1 412E Poll Control A variable that represents how process variables are brought into the transmitter for the purpose of performing a pressure compensation or other calculations. R/W Enum 128
poll_ctrl_2 412F Poll Control 2 A variable that represents how process variables are brought into the transmitter for the purpose of performing a pressure compensation or other calculations. R/W Enum 128
xmtr_tag_1 4130 External Device Tag 1 This represents the tagname for the device(s) being polled for a process variable used internally for some compensation (e.g. pressure comp). R/W PackedAscii 128
Name ID Label Description Type Access Class