Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
static_pressure | 4139 | Static Pressure | This parameter allows you to enter a static pressure when using the pressure correction feature | R/W | Float | 8 |
enable_extern_temp | 413A | Enable Ext Temp | This parameter allows you to enable/disable the external temperature feature | R/W | Enum | 8 |
static_temperature | 413B | Static Temperature | This parameter allows you to enter a static temperature when using the external temperature feature | R/W | Float | 8 |
flow_switch_setpoint | 413C | Flow Switch Setpoint | This parameter can be used to trigger a discrete output when the mass flow rate exceeds the configured setpoint. There is a fixed hysteresis of 5% for this setpoint. | R/W | Float | 16 |
flow_switch_variable_assignment | 413D | Flow Switch Variable Assignment | R/W | Enum | 8 | |
do1_assignment | 413E | DO 1 is | This parameter identifies which condition the discrete output will act upon. | R/W | Enum | 1024 |
do1_state | 413F | DO is | This parameter identifies the state of the discrete output. | R/W | Enum | 1024 |
do1_polarity | 4140 | DO 1 Polarity | This parameter allows you to set the polarity of the discrete output. | R/W | Enum | 1024 |
do1_FaultIndication | 4141 | DO Fault Indication | R/W | Enum | 1024 | |
do_index | 4142 | DO index | R/W | Index | 4096 | |
Name | ID | Label | Description | Type | Access | Class |