名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
ma_1_cutoff | 4028 | PV AO cutoff | The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. | 读/写 | 浮点 | 8 |
ma_2_cutoff | 40BC | SV AO cutoff | The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. | 读/写 | 浮点 | 8 |
sp_volume | 40BD | Base vol unit | A standard engineering unit that serves as the volume component in a special unit. | 读/写 | 枚举 | 32 |
sp_volume_time | 40BE | Base vol time | A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. | 读/写 | 枚举 | 32 |
sp_volume_fact | 40BF | Vol flo conv factr | A programmable divider for the base volume unit in a special unit. | 读/写 | 浮点 | 32 |
sp_volume_rate_text | 40C0 | Vol flo text | A 1 to 8 character description of a special unit of flow. | 读/写 | ASCII | 2048 |
sp_volume_total_text | 40C1 | Vol totl text | A 1 to 8 character description of a special unit of volume, which is based on the special unit of flow to represent total. | 读/写 | ASCII | 2048 |
fixfreq | 40C2 | 固定频率 | A value (in Hertz) that will be indicated on the frequency output for the purpose of a loop test. | 读/写 | 浮点 | 4096 |
totctl | 40C3 | 累加器控制 | Allows the user to Reset, Stop, or Start all totalizers. NOTE - Inventory totals are NOT reset. | 读/写 | 枚举 | 4096 |
fo_scaling_method | 4021 | FO Scale Method | This parameter allows you to scale the frequency output by using freq=flow, pulses/unit, or units/pulse. This parameter must be sent to the device before you make scaling changes. | 读/写 | 枚举 | 512 |