Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
cal_var_cond | 4077 | Cond | Conductivity, temperature compensated if applied | R | Float | 34 |
var_res | 4016 | Resist | Specific resistance, temperature compensated if applied | R | Float | 34 |
var_temp | 4075 | Temp | Temperature for conductivity temperature compensation if applied | R | Float | 34 |
var_salinity | 4017 | Salinity | Salinity calculated from conductivity and temperature | R | Float | 18 |
var_conductance | 4060 | Conductance Value | Conductance Value without TK | R/W | Float | 34 |
var_conc | 4061 | Concentration | Concentration Value for the selected solution | R/W | Float | 34 |
mA_output_current | 4008 | PV AO | Valor de saída analógica – valor que segue a representação do valor digital, sob modo normal de operação | R/W | Float | 66 |
percent_of_range | 400A | PV % | Porcentagem da faixa – Variável que segue a representação do valor digital em relação à faixa definida pelo valor da faixa inferior e valor da faixa superior, para modos de operação normal. As unidades dessa variável são sempre em porcentagem. | R | Float | 66 |
var_TC_lin | 4027 | Linear TC | Linear temperature coefficent used to compute the temperature compensated conductivity if applied | R/W | Float | 8 |
output_current_urv | 4033 | PV URV | Valor de faixa superior- Define o endpoint operacional do qual o valor analógico, e o ponto 100% da faixa de porcentagem são derivados. Além do que, o valor da faixa superior define um endpoint operacional no qual o alarme associado com o valor analógico, e os alarmes associados com a representação do valor digital são derivados. | R/W | Float | 64 |