Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
uiGasFactor_3 | 403E | Gas Factor | Used to make the appropriate signal sensitivity adjustment when the target gas is different from the calibration gas | R | Unsigned | 2048 |
uiActiveCounts | 403F | Active Counts | Numerical value measured by the analog-to-digital circuit monitoring the IR cell active signal | R | Unsigned | 2048 |
uiReferenceCounts | 4040 | Reference Counts | Numerical value measured by the analog-to-digital circuit monitoring the IR cell reference signal | R | Unsigned | 2048 |
uiRangeMultiplier | 4041 | RangeMultiplier | Indicates the number of decimal places in the reported concentration value. A range multiplier of 1 indicates 1 deciaml place, 10 indicates 2 decimal places, and 100 indicates 3 decimal places. | R | Unsigned | 2048 |
uiCalFactor_3 | 4042 | Cal Factor | Used to make the appropriate signal sensitivity adjustment when the calibration gas is different from the target gas | R/W | Unsigned | 2048 |
uiHeaterPower | 4043 | Heater Power | Sets the optium operating temperature for the solid state H2S sensor | R | Unsigned | 2048 |
uiHeaterVoltage | 4044 | Heater Voltage | Voltage applied to the heater film in the solid state H2S sensor | R | Unsigned | 2048 |
uiSensorResistance | 4045 | Sensor Resistance | Resistance of the sensor film in the solid state H2S sensor | R | Unsigned | 2048 |
uiHeaterCurrent | 4046 | Heater Current | Current flowing through the heater film in the solid state H2S sensor | R | Unsigned | 2048 |
uiGasFactor_5 | 4047 | Gas Factor | Used to make the appropriate signal sensitivity adjustment when the target gas is not isobutylene | R/W | Unsigned | 2048 |