名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
diaginfo_1 | 4051 | MultiEcho Dbl | 读 | 枚举 | 2050 | |
diaginfo_2 | 4052 | Tank Dim Invld | Tank Dimensions Invalid - Tank dimensions exceed maximum range of the transmitter; reduce tank height and/or Sensor Offset values. | 读 | 枚举 | 2050 |
diaginfo_3 | 4053 | Info 0008 | 读 | 枚举 | 2050 | |
diaginfo_4 | 4054 | Info 0010 | 读 | 枚举 | 2050 | |
diaginfo_5 | 4055 | Incmp Targ Rej | Incomplete Target Rejection - False target only partially captured; re-run False Target Rejection routine. | 读 | 枚举 | 2050 |
diaginfo_6 | 4056 | Scatter Target | Scattered Target - Return signals scattered due to turbulence in vessel. | 读 | 枚举 | 2050 |
diaginfo_7 | 4057 | Uncertn Target | Uncertain Target - Target is missing, but not determined to be lost; device is holding last valid value while confirming the target position, will continue based upon the LOE Delay value, consult Factory if problem persists. | 读 | 枚举 | 2050 |
diaginfo_8 | 4058 | DC Noise | DC Noise - Excessive noise on A/D channel; consult Factory. | 读 | 枚举 | 2050 |
diaginfo_9 | 4059 | Signal Too Hi | Signal Too High - Level signal amplitude is excessive; choose higher Dielectric range, reduce Turbulence or Foam settings, consult Factory if problem persists. | 读 | 枚举 | 2050 |
diaginfo_10 | 405A | Signal Too Low | Signal Too Low - Level signal amplitude is insufficient; choose lower Dielectric range, increase Turbulence or Foam settings, consult Factory if problem persists. | 读 | 枚举 | 2050 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |