Loading…
Name ID Label Description Type Access Class
diag27_text 486A Spare Indicator 7 R/W Enum 4096
diag28_text 486B Inferred Level The target has been lost within the Max Level Jump distance from the Top or Bottom Blocking Distance locations. As a result, the transmitter has inferred that the level has moved into one of those two zones, and will report level measurement of full or empty along with the Inferred Level diagnostic. Verify Level reading. If incorrect, the configuration may need to be adjusted. R/W Enum 4096
diag29_text 486C Adjust Analog Outpt Loop current is inaccurate. Perform ADJUST ANALOG OUTPUT maintenance procedure. R/W Enum 4096
diag30_text 486D Low Supply Voltage Device starved for power due to high loop resistance or inadequate voltage. Verify loop resistance Check loop power supply voltage. R/W Enum 4096
diag31_text 486E Spare Indicator 8 R/W Enum 4096
diag32_text 486F Max Jump Exceeded A potential valid level target has been detected which is further away from the last known valid level target than the distance selected for the Max Level Jump parameter. Check settings: Dielectric Range, Sensitivity. View Echo Curve. R/W Enum 4096
diag33_text 4870 Low Echo Margin Target echo has low Echo Margin rating. Check settings: Dielectric Range, Sensitivity. View Echo Curve. R/W Enum 4096
diag34_text 4871 Hi Surface Velocity The measured Surface Velocity is greater than the Max Surface Velocity value derived from the Rate of Change parameter. Confirm actual tank rate of change. Adjust (increase) Rate of Change parameter accordingly. R/W Enum 4096
diag35_text 4872 Spare Indicator 9 R/W Enum 4096
diag36_text 4873 Spare Indicator 10 R/W Enum 4096
Name ID Label Description Type Access Class