Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
diag27_text | 486A | Spare Indicator 7 | R/W | Enum | 4096 | |
diag28_text | 486B | Inferred Level | The target has been lost within the Max Level Jump distance from the Top or Bottom Blocking Distance locations. As a result, the transmitter has inferred that the level has moved into one of those two zones, and will report level measurement of full or empty along with the Inferred Level diagnostic. Verify Level reading. If incorrect, the configuration may need to be adjusted. | R/W | Enum | 4096 |
diag29_text | 486C | Adjust Analog Outpt | Loop current is inaccurate. Perform ADJUST ANALOG OUTPUT maintenance procedure. | R/W | Enum | 4096 |
diag30_text | 486D | Low Supply Voltage | Device starved for power due to high loop resistance or inadequate voltage. Verify loop resistance Check loop power supply voltage. | R/W | Enum | 4096 |
diag31_text | 486E | Spare Indicator 8 | R/W | Enum | 4096 | |
diag32_text | 486F | Max Jump Exceeded | A potential valid level target has been detected which is further away from the last known valid level target than the distance selected for the Max Level Jump parameter. Check settings: Dielectric Range, Sensitivity. View Echo Curve. | R/W | Enum | 4096 |
diag33_text | 4870 | Low Echo Margin | Target echo has low Echo Margin rating. Check settings: Dielectric Range, Sensitivity. View Echo Curve. | R/W | Enum | 4096 |
diag34_text | 4871 | Hi Surface Velocity | The measured Surface Velocity is greater than the Max Surface Velocity value derived from the Rate of Change parameter. Confirm actual tank rate of change. Adjust (increase) Rate of Change parameter accordingly. | R/W | Enum | 4096 |
diag35_text | 4872 | Spare Indicator 9 | R/W | Enum | 4096 | |
diag36_text | 4873 | Spare Indicator 10 | R/W | Enum | 4096 | |
Name | ID | Label | Description | Type | Access | Class |