Loading…
Name ID Label Description Type Access Class
vol_set4mA 40B2 Vol 4mA Set Point Defines the operational endpoint from which the Analog Value and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint from which the alarms associated with the Analog Value and the alarms associated with the Digital Value representation are derived. R/W Float 64
vol_set20mA 40B3 Vol 20mA Set Point Defines the operational endpoint from which the Analog Value and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint from which the alarms associated with the Analog Value and the alarms associated with the Digital Value representation are derived. R/W Float 64
flo_set4mA 40B4 Flow 4mA Set Point Defines the operational endpoint from which the Analog Value and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint from which the alarms associated with the Analog Value and the alarms associated with the Digital Value representation are derived. R/W Float 64
flo_set20mA 40B5 Flow 20mA Set Point Defines the operational endpoint from which the Analog Value and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint from which the alarms associated with the Analog Value and the alarms associated with the Digital Value representation are derived. R/W Float 64
DTMDescription 47C2 Description R/W ASCII 4096
ProbeModel 47C3 Probe Model Type of probe connected to the transmitter, as shown by the first four digits of the probe model number on the nameplate. Refer to the I/O Manual for additional information regarding different Probe Models. R/W Enum 2048
ProbeCoating 47C4 Probe Coating Presence of PFA coating on the probe rod. R/W Enum 2048
ProbeMount 47C5 Probe Mount Mounting style of probe process connection. R/W Enum 2048
MeasType 4097 Measurement Type The desired measurement mode of operation. (Selection of Measurement Type may be constrained by the Probe Model). R/W Enum 2048
DielRange 46A0 Dielectric Range A broad gain setting. The parameter selects the range bounding the dielectric constant of the medium being measured. For interface measurement mode, it selects the range bounding the dielectric constant of the lower liquid medium. Choose range 'Below 1.7' for very light hydrocarbons such as butane and propane. Choose range '1.7 - 3' for typical hydrocarbons and oils. Choose range '3 - 10' for amines. Choose range 'Above 10' for water-based applications. If uncertain, chose lower range setting for additional signal strength. (Some ranges may not be selectable depending on the probe model). R/W Enum 2048
Name ID Label Description Type Access Class