Name ID Label Description Type Access Class
MeasType 401B Measurement Type Selection of Measurement Type may be constrained by the probe model. R/W Enum 2048
UserPassword 401C New User Password Password required to change user parameter settings from keypad. R/W Unsigned 2048
EnteredPassword 401D Enter Password Certain parameters are password protected to prevent access by the user. Parameters that should never be changed by the user are protected with the Factory password. Some parameters intended for field use that may be changed in special, controlled situations require the Superuser password. W Unsigned 2052
FiducialType 401E Fiducial Type Fiducial Type can be used to override the default fiducial detection mode for some probe models. Default setting is suitable for most installations. Refer to I & O manual before adjusting. R/W Enum 2048
Threshold 401F Threshold Selects the negative signal threshold control for measuring the upper level pulse. Use 'CFD' except in situation with low dielectric (< 3) material over higher dielectric (> 10) material, e.g., water bottom or interface measurement, in which case use 'Fixed'. R/W Enum 2048
Window 4020 Window Factory parameter, displayed for diagnostic purposes only. R/W Unsigned 2048
Fid_Gain 4021 Fiducial Gain Default setting is suitable for most installations. Refer to I & O manual before adjusting. R/W Unsigned 2048
Sensitivity 4022 Sensitivity Sensitivity of instrument to surface of material in vessel. Default setting is suitable for most installations. Refer to I & O manual before adjusting. R/W Unsigned 2048
Trim4mA 4023 4mA Trim Value Displayed for diagnostic purposes only. R Unsigned 2049
Trim20mA 4024 20mA Trim Value Displayed for diagnostic purposes only. R Unsigned 2049